3000 ML TPN BAG Adverse Event — Malfunction (MDR 1416980-2020-02430)

3000 ML TPN BAG Adverse Event — Malfunction (MDR 1416980-2020-02430) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: 3000 ML TPN BAG; Generic name: SET, I.V. FLUID TRANSFER; Manufacturer: BAXTER HEALTHCARE CORPORATION.

Device3000 ML TPN BAG
Generic nameSET, I.V. FLUID TRANSFER
ManufacturerBAXTER HEALTHCARE CORPORATION
Report number1416980-2020-02430
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceCOMPANY REPRESENTATIVE, HEALTH
NarrativeTHE LOT WAS MANUFACTURED BETWEEN SEPTEMBER 05, 2019 TO SEPTEMBER 07, 2019. THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A FUNCTIONAL TESTING WAS PERFORMED WHICH REVEALED A LEAKAGE AT THE SPIKE PORT CAP FROM THE SPIKE PORT BASE. THE REPORTED CONDITION WAS VERIFIED. THE LIKELY CAUSE OF THE CONDITIO
SourceopenFDA MAUDE (device adverse events)