75 MM NTLC SELECTABLE RELOAD Adverse Event — Malfunction (MDR 3005075853-2020-02384)

75 MM NTLC SELECTABLE RELOAD Adverse Event — Malfunction (MDR 3005075853-2020-02384) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: 75 MM NTLC SELECTABLE RELOAD; Generic name: STAPLE, IMPLANTABLE; Manufacturer: ETHICON ENDO-SURGERY, LLC..

Device75 MM NTLC SELECTABLE RELOAD
Generic nameSTAPLE, IMPLANTABLE
ManufacturerETHICON ENDO-SURGERY, LLC.
Report number3005075853-2020-02384
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceCOMPANY REPRESENTATIVE, FOREIG
Narrative(B)(4). BATCH # T5E840. DEVICE ANALYSIS: THE ANALYSIS RESULTS FOUND THAT A SR75 RELOAD WAS RECEIVED WITH THE LEFT GRIPPING SURFACE BROKEN OFF, MAKING THE RELOAD NONFUNCTIONAL. NO FUNCTIONAL TESTING WAS PERFORMED DUE TO THE CONDITION OF THE RELOAD. THE CONDITION ON THE RETURNED RELOAD IS CONSISTENT WITH AN IMPROPER LOADING TECHNIQUE. WHEN LOADING THE CARTRIDGE IN THE DEVICE, MAKE SURE THE CARTRIDGE
SourceopenFDA MAUDE (device adverse events)