ABACUS VERSION 2 SOFTWARE Adverse Event — Injury (MDR MW5005566)
ABACUS VERSION 2 SOFTWARE Adverse Event — Injury (MDR MW5005566) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: ABACUS VERSION 2 SOFTWARE; Generic name: BAXA MICRO/MACRO COMPOUNDER; Manufacturer: BAXA.
| Device | ABACUS VERSION 2 SOFTWARE |
|---|---|
| Generic name | BAXA MICRO/MACRO COMPOUNDER |
| Manufacturer | BAXA |
| Report number | MW5005566 |
| Event type | Injury |
| Product problem | Y |
| Date received | 2008-02-18 |
| Narrative | RECEIVED CALL FROM THE HOSPITAL, ASKING IF AN ERROR IN COMPOUNDING HYDRATION COULD HAVE OCCURRED. PT'S MAGNESIUM LEVEL WAS 2.1 IN 2008, AND ON TWO DAYS LATER 9.1. THE ORDER WAS FOR D5/0.2 % SODIUM CHLORIDE WITH KPHOS 5MSQ/L + 1200 MG MGSO4/L AT 100 CC/HR CONTINUOUSLY. THE COMPOUNDER DID NOT RECOGNIZE THE MGSO4 AS A SALT AND CALCULATED THE MAGNESIUM AS AN ION. THE RESULTANT ERROR CAUSED AN OVER INF |
| Source | openFDA MAUDE (device adverse events) |
🔍 Search all device adverse events →