ALINITY C PROCESSING MODULE Adverse Event — Malfunction (MDR 3002809144-2020-00344)
ALINITY C PROCESSING MODULE Adverse Event — Malfunction (MDR 3002809144-2020-00344) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: ALINITY C PROCESSING MODULE; Generic name: AUTOMATED CHEMISTRY ANALYZER; Manufacturer: ABBOTT GMBH.
| Device | ALINITY C PROCESSING MODULE |
|---|---|
| Generic name | AUTOMATED CHEMISTRY ANALYZER |
| Manufacturer | ABBOTT GMBH |
| Report number | 3002809144-2020-00344 |
| Event type | Malfunction |
| Product problem | Y |
| Date received | 2020-04-28 |
| Report source | FOREIGN, HEALTH PROFESSIONAL |
| Narrative | THE FIELD SERVICE REPRESENTATIVE (FSR) INSPECTED THE INSTRUMENT AND REPLACED THE ALINITY C CUVETTE WASHER DRY TIP AND HIGH CONCENTRATION WASTE PERISTALTIC PUMP TUBING, WHICH RESOLVED THE CUSTOMER'S ISSUE. A REVIEW OF THE ALINITY CI PROCESSING MODULE TRACKING AND TRENDING DATA FOUND NO TRENDS ASSOCIATED WITH FALSELY DECREASED SODIUM ISSUES DESCRIBED IN THIS COMPLAINT. A REVIEW OF INSTRUMENT SERVICE |
| Source | openFDA MAUDE (device adverse events) |