← DeviceEvents
HomeDevice Adverse Events

ALINITY CI-SERIES LEVEL SENSOR, BULK SOLUTION Adverse Event — Malfunction (MDR 3002809144-2020-00347)

ALINITY CI-SERIES LEVEL SENSOR, BULK SOLUTION Adverse Event — Malfunction (MDR 3002809144-2020-00347) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: ALINITY CI-SERIES LEVEL SENSOR, BULK SOLUTION; Generic name: LEVEL SENSOR, BULK SOLUTION; Manufacturer: ABBOTT GMBH.

DeviceALINITY CI-SERIES LEVEL SENSOR, BULK SOLUTION
Generic nameLEVEL SENSOR, BULK SOLUTION
ManufacturerABBOTT GMBH
Report number3002809144-2020-00347
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceFOREIGN, HEALTH PROFESSIONAL
NarrativeCORRECTION/REMOVAL NUMBER: 3002809144-09/18/19-008-C. FURTHER INVESTIGATION INTO THIS ISSUE FOUND THAT RESULTS COULD HAVE BEEN IMPACTED BY A DEFICIENCY OF THE ALINITY CI BULK SOLUTION LEVEL SENSOR, WHERE A CRACK COULD HAVE DEVELOPED FROM ENVIRONMENTAL STRESS. A PRODUCT CORRECTION LETTER HAS BEEN ISSUED TO ALL WORLDWIDE CUSTOMERS WITH INSTALLED ALINITY I PROCESSING MODULE AND/OR ALINITY C PROCESSIN
SourceopenFDA MAUDE (device adverse events)

🔍 Search all device adverse events →