AVEA VENTILATOR Adverse Event — Injury (MDR 2021710-2020-11775)

AVEA VENTILATOR Adverse Event — Injury (MDR 2021710-2020-11775) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: AVEA VENTILATOR; Generic name: VENTILATOR, CONTINUOUS, FACILITY USE; Manufacturer: VYAIRE MEDICAL.

DeviceAVEA VENTILATOR
Generic nameVENTILATOR, CONTINUOUS, FACILITY USE
ManufacturerVYAIRE MEDICAL
Report number2021710-2020-11775
Event typeInjury
Product problemY
Date received2020-04-27
Report sourceHEALTH PROFESSIONAL, USER FACI
NarrativeAN ONSITE PERSONNEL FOUND OUT THAT THERE WAS A LEAK IN EXHALATION CORNER. HE TOOK IT APART AND RESET IT ALL WITH 2% LEAK VOLUMES BACK TO NORMAL. HE ALSO FIXED THE PEEP BY TURNING EXH VALVE. A TEST WAS CONDUCTED AND THE DEVICE WAS RUNNING WELL AS IT SHOULD ACCORDING TO SPECIFICATION. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION