CADD LEGACY PUMP Adverse Event — Malfunction (MDR 3012307300-2020-03380)
CADD LEGACY PUMP Adverse Event — Malfunction (MDR 3012307300-2020-03380) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: CADD LEGACY PUMP; Generic name: PUMP, INFUSION; Manufacturer: ST PAUL.
| Device | CADD LEGACY PUMP |
|---|---|
| Generic name | PUMP, INFUSION |
| Manufacturer | ST PAUL |
| Report number | 3012307300-2020-03380 |
| Event type | Malfunction |
| Product problem | Y |
| Date received | 2020-04-27 |
| Report source | DISTRIBUTOR |
| Narrative | ONE SMITHS MEDICAL CADD LEGACY PUMP WAS RETURNED FOR ANALYSIS WITH NO PHYSICAL DAMAGE NOTED ON THE DEVICE. EVENT LOG HISTORY WAS PERFORMED AND INDICATED THAT ERROR 1871 WAS RECORDED IN HISTORY. DURING ANALYSIS, THE REPORTED PROBLEM REGARDING ERROR 1871 WAS ABLE TO BE DUPLICATED. IT WAS NOTED THAT THE MOTOR WAS FAULTY AND WAS THE CAUSE OF THE REPORTED ISSUE. SERVICE REPLACED THE MOTOR TO CORRECT TH |
| Source | openFDA MAUDE (device adverse events) |
🔍 Search all device adverse events →