CARELINK PROGRAMMER Adverse Event — Malfunction (MDR 2182208-2020-00822) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: CARELINK PROGRAMMER; Generic name: PROGRAMMER, PACEMAKER; Manufacturer: MEDTRONIC, INC..

DeviceCARELINK PROGRAMMER
Generic namePROGRAMMER, PACEMAKER
ManufacturerMEDTRONIC, INC.
Report number2182208-2020-00822
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceFOREIGN, HEALTH PROFESSIONAL
NarrativePRODUCT ANALYSIS: ANALYSIS WAS UNABLE TO CONFIRM THAT THE PROGRAMMER WOULD NOT POWER UP NOTING THAT THE PROGRAMMER POWERED UP AS NORMAL. ANALYSIS WAS ABLE TO CONFIRM THAT THE CURSOR AND THE STYLUS WERE MISALIGNED. ANALYSIS ALSO FOUND THAT A SCREW FROM THE XY BOARD FELL OUT AND WAS RATTLING IN THE DISPLAY AREA, THE STYLUS WAS MISSING, AND THE HANDLE COVERS WERE CRACKED. ALL FOUND DEFECTIVE PARTS WE
SourceopenFDA MAUDE (device adverse events)