← DeviceEvents
HomeDevice Adverse Events

CM ALVIM ACQUA IMPLANT 3.5X13 MM Adverse Event — Injury (MDR 3008261720-2020-02119)

CM ALVIM ACQUA IMPLANT 3.5X13 MM Adverse Event — Injury (MDR 3008261720-2020-02119) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: CM ALVIM ACQUA IMPLANT 3.5X13 MM; Generic name: ENDOSSEOUS DENTAL IMPLANT; Manufacturer: NEODENT - JJGC S.A..

DeviceCM ALVIM ACQUA IMPLANT 3.5X13 MM
Generic nameENDOSSEOUS DENTAL IMPLANT
ManufacturerNEODENT - JJGC S.A.
Report number3008261720-2020-02119
Event typeInjury
Product problemN
Date received2020-04-27
Report sourceFOREIGN, HEALTH PROFESSIONAL
NarrativeCONSIDERING THE SURGICAL METHODOLOGY, IT WAS SEEN THAT THE DENTIST HAS USED SEQUENCE OF DRILLS DIFFERENT THAN THE ONE RECOMMENDED FOR THE IMPLANT INSTALLATION. THE INVESTIGATION OF THE COMPLAINT WAS CARRIED OUT CONSIDERING THE ANALYSIS OF THE INFORMATION GIVEN BY THE DENTIST IN THE GUARANTEE FORM, VISUAL CONFERENCE OF THE PRODUCT RETURNED BY THE DENTIST AND THE EVALUATION OF RELEVANT PRODUCTION RE
SourceopenFDA MAUDE (device adverse events)

🔍 Search all device adverse events →