CPS TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE 3-5 EF TOP Adverse Event — Malfunction (MDR 0001822565-2020-01535)

CPS TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE 3-5 EF TOP Adverse Event — Malfunction (MDR 0001822565-2020-01535) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: CPS TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE 3-5 EF TOP; Generic name: PROSTHESIS, KNEE; Manufacturer: ZI

DeviceCPS TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE 3-5 EF TOP
Generic namePROSTHESIS, KNEE
ManufacturerZIMMER BIOMET, INC.
Report number0001822565-2020-01535
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceCOMPANY REPRESENTATIVE, DISTRI
NarrativeTHIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4; B5; G4; G7; H1; H2; H3; H6 COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL EXAMINATION EXHIBITS SIGNS OF REPEATED USE AND THE ANTERIOR OF THE POST FEATURE HAS FRACTURED OFF. DEVICE HISTORY RECORD (DHR) REVIEWED FOR DEVIATIONS AND/ OR ANOMALIES WITH NO DEVIA
SourceopenFDA MAUDE (device adverse events)