DIMENSION CLINICAL CHEMISTRY SYSTEM Adverse Event — Malfunction (MDR 1226181-2008-00013)

DIMENSION CLINICAL CHEMISTRY SYSTEM Adverse Event — Malfunction (MDR 1226181-2008-00013) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: DIMENSION CLINICAL CHEMISTRY SYSTEM; Generic name: CLINICAL CHEMISTRY SYSTEM; Manufacturer: SIEMENS HEALTHCARE DIAGNOSTICS INC..

DeviceDIMENSION CLINICAL CHEMISTRY SYSTEM
Generic nameCLINICAL CHEMISTRY SYSTEM
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Report number1226181-2008-00013
Event typeMalfunction
Product problemY
Date received2008-02-19
Report sourceHealth Professional
NarrativeA FALSELY ELEVATED TROPONIN I RESULT OF 19.33 NG/ML WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS RETESTED AND A RESULT OF <0.04 NG/ML WAS OBTAINED. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT. ANALYSIS OF THE INSTRUMEN