DVR ANATOMIC EXT LEFT Adverse Event — Injury (MDR 0001825034-2020-01796)

DVR ANATOMIC EXT LEFT Adverse Event — Injury (MDR 0001825034-2020-01796) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: DVR ANATOMIC EXT LEFT; Generic name: PLATE, FIXATION; Manufacturer: ZIMMER BIOMET, INC..

DeviceDVR ANATOMIC EXT LEFT
Generic namePLATE, FIXATION
ManufacturerZIMMER BIOMET, INC.
Report number0001825034-2020-01796
Event typeInjury
Product problemY
Date received2020-04-28
Report sourceCOMPANY REPRESENTATIVE, FOREIG
NarrativeTHIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. VISUAL EXAMINATION OF THE PROVIDED PICTURES IDENTIFIED THE PLATE IS FRACTURED AT THE SCREW HOLES. X-RAYS PROVIDED WERE SENT TO THE HCP FOR REVIEW. THEIR REVIEW NOTED THE OVERALL FIT OF THE IMPLANTS APPEARS NORMAL. AS A RESULT OF THE PLATE FRACTURE, THERE IS ABNORMAL ALIGNMENT WITH APEX VOLAR ANGULATION OF THE BONE AND PLATE.