ENDOWRIST Adverse Event — Malfunction (MDR 2955842-2020-10346)

ENDOWRIST Adverse Event — Malfunction (MDR 2955842-2020-10346) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: ENDOWRIST; Generic name: SMALL GRASPING RETRACTOR; Manufacturer: INTUITIVE SURGICAL, INC.

DeviceENDOWRIST
Generic nameSMALL GRASPING RETRACTOR
ManufacturerINTUITIVE SURGICAL, INC
Report number2955842-2020-10346
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceOTHER
NarrativeINTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE SMALL GRASPING RETRACTOR ASSOCIATED WITH THIS COMPLAINT AND COMPLETED INVESTIGATIONS. FAILURE ANALYSIS INVESTIGATIONS REPLICATED AND CONFIRMED THE CUSTOMER REPORTED COMPLAINT. THE INSTRUMENT WAS FOUND TO HAVE A BROKEN PITCH CABLE AT THE DISTAL END. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. A SITE HISTORY REVI