ENDOWRIST Adverse Event — Malfunction (MDR 2955842-2020-10369)

ENDOWRIST Adverse Event — Malfunction (MDR 2955842-2020-10369) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: ENDOWRIST; Generic name: PERMANENT CAUTERY HOOK; Manufacturer: INTUITIVE SURGICAL, INC.

DeviceENDOWRIST
Generic namePERMANENT CAUTERY HOOK
ManufacturerINTUITIVE SURGICAL, INC
Report number2955842-2020-10369
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceOTHER
NarrativeINTUITIVE SURGICAL, INC. (ISI) RECEIVED THE PERMANENT CAUTERY HOOK INSTRUMENT AND PERFORMED A DEVICE EVALUATION. FAILURE ANALYSIS INVESTIGATIONS CONFIRMED THE CUSTOMER REPORTED COMPLAINT OF "ARCING ISSUE." THE INSTRUMENT WAS FOUND TO HAVE THERMAL DAMAGE ON THE MONOPOLAR YAW PULLEY. THE INSTRUMENT WAS FOUND TO HAVE DAMAGE OF THE CONDUCTOR WIRE¿S INSULATION. THE DISTAL CLEVIS EAR WAS CUT TO INSPECT