← DeviceEvents
HomeDevice Adverse Events

KIT MMT-1755K PUMP 630G 3ML BK MG Adverse Event — Malfunction (MDR 2032227-2020-110418)

KIT MMT-1755K PUMP 630G 3ML BK MG Adverse Event — Malfunction (MDR 2032227-2020-110418) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: KIT MMT-1755K PUMP 630G 3ML BK MG; Generic name: ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND; Manufacturer: MEDTRONIC PUERTO RICO OPER

DeviceKIT MMT-1755K PUMP 630G 3ML BK MG
Generic nameARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Report number2032227-2020-110418
Event typeMalfunction
Product problemY
Date received2020-04-27
Report sourceForeign, Company representation
NarrativeDEVICE RECEIVED WITH MISSING RETAINER. UNABLE TO PERFORM THE DISPLACEMENT TEST AND P-CAP / RESERVOIR WILL NOT LOCK PROPERLY DUE TO MISSING RETAINER. DEVICE RECEIVED WITH SCRATCHED CASE. DEVICE RECEIVED WITH PILLOWING KEYPAD OVERLAY.
SourceopenFDA MAUDE (device adverse events)

🔍 Search all device adverse events →