← DeviceEvents
HomeDevice Adverse Events

LEIBINGER Adverse Event — Malfunction (MDR MW5005570)

LEIBINGER Adverse Event — Malfunction (MDR MW5005570) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: LEIBINGER; Generic name: 2.0MM SELF-TAPPING SCREW; Manufacturer: LEIBINGER.

DeviceLEIBINGER
Generic name2.0MM SELF-TAPPING SCREW
ManufacturerLEIBINGER
Report numberMW5005570
Event typeMalfunction
Product problemY
Date received2008-01-18
NarrativePREOPERATIVE DIAGNOSIS OF MANDIBULAR HYPOPLASIA. ADMITTED IN 2008 TO THE SAME DAY SURGERY CENTER AND UNDERWENT A BILATERAL SAGITTAL SPLIT OSTEOTOMY WITH MIDLINE OSTEOTOMY. INTRAOPERATIVELY THE 2.0MM SELF-TAPPING SCREW FRACTURED UNDERNEATH THE HEAD. -TORSIONAL FAILURE- THE THREADED BOTTOM OF THE SCREW REMAINS IN THE PATIENT MINUS THE SCREW HEAD. NO PATIENT HARM AS PER THE SURGEON.
SourceopenFDA MAUDE (device adverse events)