← DeviceEvents
HomeDevice Adverse Events

LEVEL 1 HOTLINE FLUID WARMING ACCESSORY Adverse Event — Malfunction (MDR 3012307300-2020-03434)

LEVEL 1 HOTLINE FLUID WARMING ACCESSORY Adverse Event — Malfunction (MDR 3012307300-2020-03434) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: LEVEL 1 HOTLINE FLUID WARMING ACCESSORY; Generic name: WARMER, THERMAL, INFUSION FLUID; Manufacturer: ST PAUL.

DeviceLEVEL 1 HOTLINE FLUID WARMING ACCESSORY
Generic nameWARMER, THERMAL, INFUSION FLUID
ManufacturerST PAUL
Report number3012307300-2020-03434
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceUSER FACILITY
NarrativeEVALUATION RESULTS: ONE HL-90 115V ELECTRONIC ASSEMBLY (LEVEL 1 HOTLINE FLUID WARMING ACCESSORY) WAS RETURNED FOR INVESTIGATION IN USED CONDITION. VISUAL INSPECTION REVEALED THAT PCB BOARD WAS IN GOOD CONDITION. THE INVESTIGATOR FIRST INSTALLED THE PCB BOARD TO A TEST UNIT. THE INVESTIGATOR THEN FILLED THE TANK WITH WATER, PLUGGED IN THE LINE CORD, AND TURNED ON THE POWER SWITCH. THE CUSTOMER REPO
SourceopenFDA MAUDE (device adverse events)

🔍 Search all device adverse events →