← DeviceEvents
HomeDevice Adverse Events

LOCKING TITANIUM ADAPTER Adverse Event — Malfunction (MDR 1416980-2020-02398)

LOCKING TITANIUM ADAPTER Adverse Event — Malfunction (MDR 1416980-2020-02398) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: LOCKING TITANIUM ADAPTER; Generic name: SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE; Manufacturer: BAXTER HEALTHCARE CORPORATION.

DeviceLOCKING TITANIUM ADAPTER
Generic nameSET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
ManufacturerBAXTER HEALTHCARE CORPORATION
Report number1416980-2020-02398
Event typeMalfunction
Product problemY
Date received2020-04-27
Report sourceCOMPANY REPRESENTATIVE, FOREIG
NarrativeADDITIONAL INFORMATION: D10, H3, AND H6. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A CONNECTION TEST AND DIMENSIONAL INSPECTION WERE PERFORMED WITH NO ISSUES; ALL SPECIFICATIONS WERE MET. THE DEVICE WAS FOUND TO BE CONFORMING PRODUCT. THE REPORTED CONDITION WAS NOT VERIFIED. SHOULD ADDI
SourceopenFDA MAUDE (device adverse events)