PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWH Adverse Event — Malfunction (MDR 2032227-2020-110315)
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWH Adverse Event — Malfunction (MDR 2032227-2020-110315) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWH; Generic name: PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE
| Device | PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWH |
|---|---|
| Generic name | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SE |
| Manufacturer | MEDTRONIC PUERTO RICO OPERATIONS CO. |
| Report number | 2032227-2020-110315 |
| Event type | Malfunction |
| Product problem | Y |
| Date received | 2020-04-27 |
| Report source | COMPANY REPRESENTATIVE, FOREIG |
| Narrative | DEVICE PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME OR A33 AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED. NO E70 - MOTOR POSITION ENCODER ERROR ALARM NOTED. MOTOR PASSED MOTOR TEST. (B)(4). |
| Source | openFDA MAUDE (device adverse events) |