← DeviceEvents
HomeDevice Adverse Events

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB Adverse Event — Malfunction (MDR 2032227-2020-111084)

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB Adverse Event — Malfunction (MDR 2032227-2020-111084) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB; Generic name: PUMP, INFUSION, INSULIN, TO BE USED WITH INV

DevicePARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB
Generic namePUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SE
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Report number2032227-2020-111084
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceCONSUMER
NarrativePASS DISPLACEMENT TEST. DEVICE RECEIVED A33 ALARM DURING THE BASIC OCCLUSION TEST DUE TO LOOSE OR PROTRUDED DRIVE SUPPORT DISK. UNABLE TO PERFORM PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST AND OCCLUSION TEST DUE TO A33 ALARM.
SourceopenFDA MAUDE (device adverse events)

🔍 Search all device adverse events →