PERMANENT PACEMAKER ELECTRODE Adverse Event — Malfunction (MDR 2938836-2020-02846)

PERMANENT PACEMAKER ELECTRODE Adverse Event — Malfunction (MDR 2938836-2020-02846) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PERMANENT PACEMAKER ELECTRODE; Generic name: PERMANENT PACEMAKER ELECTRODE; Manufacturer: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE).

DevicePERMANENT PACEMAKER ELECTRODE
Generic namePERMANENT PACEMAKER ELECTRODE
ManufacturerST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Report number2938836-2020-02846
Event typeMalfunction
Product problemY
Date received2020-04-27
Report sourceOTHER
NarrativeNO COMPLAINT WAS RECEIVED WITH THE RETURN OF THE DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. A PARTIAL LEAD WITH THE DISTAL TIP MEASURING 41.5 / 44.0 / 46.5 CM INNER AND OUTER INSULATIONS /OUTER COIL / INNER COIL. FINAL ANALYSIS FOUND CLAVICULAR CRUSH DAMAGE DISTAL TO THE SUTURE SLEEVE TIE IMPRESSION DAMAGING THE OUTER INSULATION. OTHER DAMAGES FOUND ARE CONSISTENT WITH PROCEDURAL DAMAGE.