← DeviceEvents
HomeDevice Adverse Events

PERSONA CPS TIBIAL ARTICULAR SURFACE PROVISIONAL Adverse Event — Malfunction (MDR 0001822565-2020-01562)

PERSONA CPS TIBIAL ARTICULAR SURFACE PROVISIONAL Adverse Event — Malfunction (MDR 0001822565-2020-01562) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PERSONA CPS TIBIAL ARTICULAR SURFACE PROVISIONAL; Generic name: PROSTHESIS, KNEE; Manufacturer: ZIMMER BIOMET, INC..

DevicePERSONA CPS TIBIAL ARTICULAR SURFACE PROVISIONAL
Generic namePROSTHESIS, KNEE
ManufacturerZIMMER BIOMET, INC.
Report number0001822565-2020-01562
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceDISTRIBUTOR
NarrativeTHIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL EVALUATION OF THE RETURNED DEVICE SHOWS SIGNS OF REPEATED USE AND IS FRACTURED ON MEDIAL SIDE OF POST. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. EVALUATION OF THE RETURNED DEVICE IDENTIFI
SourceopenFDA MAUDE (device adverse events)