← DeviceEvents
HomeDevice Adverse Events

PLEXA S 60 Adverse Event — Injury (MDR 1028232-2020-01873)

PLEXA S 60 Adverse Event — Injury (MDR 1028232-2020-01873) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PLEXA S 60; Generic name: ICD LEAD; Manufacturer: BIOTRONIK SE & CO. KG.

DevicePLEXA S 60
Generic nameICD LEAD
ManufacturerBIOTRONIK SE & CO. KG
Report number1028232-2020-01873
Event typeInjury
Product problemN
Date received2020-04-28
Report sourceCOMPANY REPRESENTATIVE
NarrativeTHE METHOD, RESULT, AND CONCLUSION CODES REMAIN THE SAME AS FOLLOW UP 01. AN INFECTION WAS OBSERVED FOLLOWING THE IMPLANTATION OF THIS BIOTRONIK DEVICE. THE STERILIZATION PROCESS WAS INVESTIGATED. THE VALIDATED PROCESS ASSURES THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., ARE WITHIN ITS SPECIFIED RANGES FOR EACH DISTRIBUTED DEVICE. ADDITIONALLY AN ANALY
SourceopenFDA MAUDE (device adverse events)