POWERED PLUS LONG 45 Adverse Event — Malfunction (MDR 3005075853-2020-02345) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: POWERED PLUS LONG 45; Generic name: STAPLE, IMPLANTABLE; Manufacturer: ETHICON ENDO-SURGERY, LLC..

DevicePOWERED PLUS LONG 45
Generic nameSTAPLE, IMPLANTABLE
ManufacturerETHICON ENDO-SURGERY, LLC.
Report number3005075853-2020-02345
Event typeMalfunction
Product problemY
Date received2020-04-27
Report sourceHEALTH PROFESSIONAL
Narrative(B)(4). DATE SENT: 5/18/2020. D4: BATCH # P5604E. DEVICE ANALYSIS: THE ANALYSIS FOUND THAT ONE PLE45A DEVICE WAS RETURNED WITH NO APPARENT DAMAGE AND WITH NO CARTRIDGE LOADED ON THE DEVICE. THE MANUAL OVERRIDE DOOR WAS OUT OF POSITION AND MISSING. THE OVERRIDE LEVER WAS UP, WHICH DENOTES THAT THE KNIFE WAS MANUALLY RETURNED TO HOME POSITION. IT SHOULD BE NOTED AFTER THE MANUAL OVERRIDE SYSTEM IS U