← DeviceEvents
HomeDevice Adverse Events

PUMP MMT-1780KPK 670G PATHWAY BLACK MG Adverse Event — Malfunction (MDR 2032227-2020-111161)

PUMP MMT-1780KPK 670G PATHWAY BLACK MG Adverse Event — Malfunction (MDR 2032227-2020-111161) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PUMP MMT-1780KPK 670G PATHWAY BLACK MG; Generic name: ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL; Manufacturer: MEDTRONIC

DevicePUMP MMT-1780KPK 670G PATHWAY BLACK MG
Generic nameARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Report number2032227-2020-111161
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceCONSUMER
NarrativeDEVICE RECEIVED WITH BLANK DISPLAY DUE TO CORRODED ELECTRONICS ASSEMBLY. MOTOR AND FORCE SENSOR WAS ALSO CORRODED. UNABLE TO PERFORM DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO BLANK DISPLAY. UNABLE TO VERIFY CRITICAL PUMP ERROR OR OPEN BOOK IMAGE DUE TO BLANK DISPLAY.
SourceopenFDA MAUDE (device adverse events)

🔍 Search all device adverse events →