PUMP MMT-1780KPK 670G PATHWAY BLACK MG Adverse Event — Malfunction (MDR 2032227-2020-111161)
PUMP MMT-1780KPK 670G PATHWAY BLACK MG Adverse Event — Malfunction (MDR 2032227-2020-111161) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PUMP MMT-1780KPK 670G PATHWAY BLACK MG; Generic name: ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL; Manufacturer: MEDTRONIC
| Device | PUMP MMT-1780KPK 670G PATHWAY BLACK MG |
|---|---|
| Generic name | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL |
| Manufacturer | MEDTRONIC PUERTO RICO OPERATIONS CO. |
| Report number | 2032227-2020-111161 |
| Event type | Malfunction |
| Product problem | Y |
| Date received | 2020-04-28 |
| Report source | CONSUMER |
| Narrative | DEVICE RECEIVED WITH BLANK DISPLAY DUE TO CORRODED ELECTRONICS ASSEMBLY. MOTOR AND FORCE SENSOR WAS ALSO CORRODED. UNABLE TO PERFORM DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO BLANK DISPLAY. UNABLE TO VERIFY CRITICAL PUMP ERROR OR OPEN BOOK IMAGE DUE TO BLANK DISPLAY. |