← DeviceEvents
HomeDevice Adverse Events

PWRD ECH FLEX 60MM 34CM SHAFT Adverse Event — Malfunction (MDR 3005075853-2020-02354)

PWRD ECH FLEX 60MM 34CM SHAFT Adverse Event — Malfunction (MDR 3005075853-2020-02354) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: PWRD ECH FLEX 60MM 34CM SHAFT; Generic name: STAPLE, IMPLANTABLE; Manufacturer: ETHICON ENDO-SURGERY, LLC..

DevicePWRD ECH FLEX 60MM 34CM SHAFT
Generic nameSTAPLE, IMPLANTABLE
ManufacturerETHICON ENDO-SURGERY, LLC.
Report number3005075853-2020-02354
Event typeMalfunction
Product problemY
Date received2020-04-27
Report sourceCOMPANY REPRESENTATIVE, FOREIG
Narrative(B)(4). DATE SENT: 5/18/2020. D4: BATCH # T5EK3W. DEVICE ANALYSIS: THE ANALYSIS FOUND THAT ONE PSE60A DEVICE WAS RETURNED WITH NO APPARENT DAMAGE AND WITH AN ECR60B CARTRIDGE RELOAD LOADED ON THE DEVICE. THE CARTRIDGE WAS RECEIVED PARTIALLY FIRED 1/16 WITH ONE PIECE SLED DAMAGED. IN ADDITION, ANOTHER ECR60D RELOAD (B) WAS RECEIVED FULLY FIRED WITH FOUR DOUBLE DRIVERS MISSING AND ONE DOUBLE DRIVER
SourceopenFDA MAUDE (device adverse events)