← DeviceEvents
HomeDevice Adverse Events

SELECTSECURE MRI SURESCAN Adverse Event — Injury (MDR 2649622-2020-08359)

SELECTSECURE MRI SURESCAN Adverse Event — Injury (MDR 2649622-2020-08359) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: SELECTSECURE MRI SURESCAN; Generic name: DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIA; Manufacturer: MPRI.

DeviceSELECTSECURE MRI SURESCAN
Generic nameDRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIA
ManufacturerMPRI
Report number2649622-2020-08359
Event typeInjury
Product problemY
Date received2020-04-27
Report sourceHEALTH PROFESSIONAL
NarrativePRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND NO ANOMALIES WERE FOUND. THE DISTAL LOW VOLTAGE ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE
SourceopenFDA MAUDE (device adverse events)

🔍 Search all device adverse events →