SERVO-I VENTILATOR Adverse Event — Malfunction (MDR MW5005585)
SERVO-I VENTILATOR Adverse Event — Malfunction (MDR MW5005585) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: SERVO-I VENTILATOR; Generic name: SAME; Manufacturer: MAQUET.
| Device | SERVO-I VENTILATOR |
|---|---|
| Generic name | SAME |
| Manufacturer | MAQUET |
| Report number | MW5005585 |
| Event type | Malfunction |
| Product problem | Y |
| Date received | 2008-02-15 |
| Narrative | AFTER EVALUATION BY OUR BIO-MEDICAL STAFF, IT WAS DISCOVERED THAT THERE WAS WATER INTRUSION INTO THE PRESSURE TUBING FROM THE EXHALATION CARTRIDGE TO THE PRESSURE TRANSDUCER ON THE VENTILATOR CAUSING VARIATIONS OF "SET PEEP" AND "MEASURED PEEP" ON THE SERVO-I VENTILATOR. |
| Source | openFDA MAUDE (device adverse events) |