SERVO-I VENTILATOR Adverse Event — Malfunction (MDR MW5005585)

SERVO-I VENTILATOR Adverse Event — Malfunction (MDR MW5005585) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: SERVO-I VENTILATOR; Generic name: SAME; Manufacturer: MAQUET.

DeviceSERVO-I VENTILATOR
Generic nameSAME
ManufacturerMAQUET
Report numberMW5005585
Event typeMalfunction
Product problemY
Date received2008-02-15
NarrativeAFTER EVALUATION BY OUR BIO-MEDICAL STAFF, IT WAS DISCOVERED THAT THERE WAS WATER INTRUSION INTO THE PRESSURE TUBING FROM THE EXHALATION CARTRIDGE TO THE PRESSURE TRANSDUCER ON THE VENTILATOR CAUSING VARIATIONS OF "SET PEEP" AND "MEASURED PEEP" ON THE SERVO-I VENTILATOR.
SourceopenFDA MAUDE (device adverse events)