← DeviceEvents
HomeDevice Adverse Events

SOLERO MICROWAVE TISSUE ABLATION APPLICATOR (14CM) Adverse Event — Malfunction (MDR 1317056-2020-00075)

SOLERO MICROWAVE TISSUE ABLATION APPLICATOR (14CM) Adverse Event — Malfunction (MDR 1317056-2020-00075) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: SOLERO MICROWAVE TISSUE ABLATION APPLICATOR (14CM); Generic name: SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES; Manufacturer: AN

DeviceSOLERO MICROWAVE TISSUE ABLATION APPLICATOR (14CM)
Generic nameSYSTEM, ABLATION, MICROWAVE AND ACCESSORIES
ManufacturerANGIODYNAMICS
Report number1317056-2020-00075
Event typeMalfunction
Product problemY
Date received2020-04-28
Report sourceCOMPANY REPRESENTATIVE, HEALTH
NarrativeRETURNED FOR EVALUATION WAS A SOLERO PROBE. THE RETURNED PROBE WAS REVIEWED BY R&D AND CONFIRMED THE CERAMIC TIP WAS FRACTURED AT THE END OF THE SEMI-RIGID SLEEVE. THE ROOT CAUSE OF THE FRACTURE WAS A THERMAL EVENT THAT OVER-HEATED THE SEMI-RIGID CAUSING THE CERAMIC TIP TO FRACTURE. THE ROOT CAUSE OF THE THERMAL EVENT CANNOT BE DEFINITELY DETERMINED BUT IS POTENTIALLY RELATED TO A VARIATION IN THE
SourceopenFDA MAUDE (device adverse events)