THERMOSONIC Adverse Event — Malfunction (MDR 10005092)
THERMOSONIC Adverse Event — Malfunction (MDR 10005092) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: THERMOSONIC; Generic name: MEDIA, COUPLING, ULTRASOUND; Manufacturer: PARKER LABORATORIES, INC..
| Device | THERMOSONIC |
|---|---|
| Generic name | MEDIA, COUPLING, ULTRASOUND |
| Manufacturer | PARKER LABORATORIES, INC. |
| Report number | 10005092 |
| Event type | Malfunction |
| Product problem | Y |
| Date received | 2020-04-28 |
| Narrative | AQUASONIC ULTRASOUND TRANSMISSION GEL WARMER MACHINE HAD WARMED THE GEL TOO HIGH WHERE THE NURSE'S LEFT HAND WAS BURNT WHEN USING IT. NO GEL CAME IN CONTACT WITH THE PATIENT. THE RN REQUIRED AN ICE PACK AFTER THE INCIDENT BUT DENIED ADDITIONAL MEDICAL TREATMENT. NO BLISTERING WAS REPORTED. THE WARMING UNIT WAS SEQUESTERED BY BIOMEDICAL ENGINEERING AND WAS TESTED. IT WAS HEATING TO 153 DEGREES F AN |
| Source | openFDA MAUDE (device adverse events) |