DeviceTRIMA ACCEL
Generic nameTRIMA ACCEL RBC PLT PLASMA SET
ManufacturerTERUMO BCT
Report number1722028-2020-00188
Event typeMalfunction
Product problemY
Date received2020-04-27
Report sourceHEALTH PROFESSIONAL
NarrativeINVESTIGATION: THE RUN DATA FILE (RDF) WAS ANALYZED FOR THIS EVENT. THE SIGNALS IN THE RUN DATA FILE INDICATE THAT THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT WAS LIKELY A RESULT OF AN ESCAPE OF WBC FROM THE LRS CHAMBER DURING THE PROCEDURE. BASED ON THE AVAILABLE INFORMATION, IT CANNOT BE RULED OUT THAT THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT COULD BE DONOR-R
SourceopenFDA MAUDE (device adverse events)