UL LIBERTY CYCLER ASSEMBLY(NON-VALUATED) Adverse Event — Malfunction (MDR 2937457-2020-00750)

UL LIBERTY CYCLER ASSEMBLY(NON-VALUATED) Adverse Event — Malfunction (MDR 2937457-2020-00750) is one of 9,000 device adverse events in the DeviceEvents dataset. Key details — Device: UL LIBERTY CYCLER ASSEMBLY(NON-VALUATED); Generic name: SYSTEM, PERITONEAL, AUTOMATIC DELIVERY; Manufacturer: CONCORD MANUFACTURING.

DeviceUL LIBERTY CYCLER ASSEMBLY(NON-VALUATED)
Generic nameSYSTEM, PERITONEAL, AUTOMATIC DELIVERY
ManufacturerCONCORD MANUFACTURING
Report number2937457-2020-00750
Event typeMalfunction
Product problemY
Date received2020-04-27
Report sourceCONSUMER
NarrativePLANT INVESTIGATION: THE ACTUAL DEVICE WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. AN EXTERIOR VISUAL INSPECTION OF THE RETURNED CYCLER SHOWED NO SIGNS OF PHYSICAL DAMAGE. UPON POWER UP, THE CYCLER TOUCH SCREEN TEST FAILED. WHEN POWERING ON THE CYCLER, THE OK, STOP AND UP/DOWN ARROW PUSH BUTTONS ILLUMINATED, HOWEVER THE FRONT PANEL TOUCH SCREEN REMAINED BLANK. IT WAS IDENTIFIED THAT