Abbott Laboratories, Inc Recall: ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part n

Abbott Laboratories, Inc Recall: ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part n is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Abbott Laboratories, Inc; Product: ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part number 01G06; ARCHITECT c16000 part number 03L77 Clinica

Recalling firmAbbott Laboratories, Inc
ProductARCHITECT c4000 part number 02P24; ARCHITECT c8000 part number 01G06; ARCHITECT c16000 part number 03L77 Clinical Chemistry systems.
Reason for recallThe Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusCompleted
DistributionWorldwide distribution to US, Canada, France, and Germany.
Recall initiated20170929
Recall numberZ-0165-2018
Categorydevice
StateTX
CountryUnited States