Abbott Laboratories, Inc Recall: Clinical Chemistry Phenobarbital, list number 1E08
Abbott Laboratories, Inc Recall: Clinical Chemistry Phenobarbital, list number 1E08 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Abbott Laboratories, Inc; Product: Clinical Chemistry Phenobarbital, list number 1E08; Reason for recall: Abbott Diagnostics is recalling Clinical Chemist
| Recalling firm | Abbott Laboratories, Inc |
|---|---|
| Product | Clinical Chemistry Phenobarbital, list number 1E08 |
| Reason for recall | Abbott Diagnostics is recalling Clinical Chemistry Phenobarbital reagent lots 62299UN12 and 85773UN12 due to imprecision issues which increase with the age of the reagent lot. The imprecision of the Phenobarbital assay is d 7% Total CV%; however, internal testing confirmed CVs up to approximately 12% for in-date reagent lots over five months old. Although lot 52803UN12 is now expired, it exhibit |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally to Internationally distributed to Argentina, Australia, Austria, Bahamas, Brunei, Canada, Chile, Curacao, Denmark, France, Germany, Greece, Ireland, Italy, Lebanon, Mexico, Netherlands, Norway, Oman, Paraguay, Portugal, Puerto Rico, Russi |
| Recall initiated | 20121116 |
| Recall number | Z-1118-2013 |
| Category | device |
| State | TX |
| Country | United States |