← RecallGuard
HomeRecalls

Abbott Medical Recall: Infinity 7 Implantable Pulse Generator REF 6662 Product Description

Abbott Medical Recall: Infinity 7 Implantable Pulse Generator REF 6662 Product Description is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Abbott Medical; Product: Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electroni

Recalling firmAbbott Medical
ProductInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system program
Reason for recallFirm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusOngoing
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The count
Recall initiated20230718
Recall numberZ-2511-2023
Categorydevice
StateTX
CountryUnited States

🔍 Search all recalls →