Abbott Recall: Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060
Abbott Recall: Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Abbott; Product: Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 and 9-TVLP4F90/080; Reason for recall: Device may have a small b
| Recalling firm | Abbott |
|---|---|
| Product | Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 and 9-TVLP4F90/080 |
| Reason for recall | Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide - US Nationwide distribution. |
| Recall initiated | 20250317 |
| Recall number | Z-1643-2025 |
| Category | device |
| State | MN |
| Country | United States |