Accuray Incorporated Recall: RoboCouch Patient Positioning System Product Usage: is intend is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Accuray Incorporated; Product: RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient
| Recalling firm | Accuray Incorporated |
|---|---|
| Product | RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required. |
| Reason for recall | A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution - USA Nationwide including the states of: OH, MO. OK, CA and VA and the countries of: Finland, France, Luxembourg, Poland, Russia, Turkey and Japan. |
| Recall initiated | 20130327 |
| Recall number | Z-1155-2013 |
| Category | device |
| State | CA |
| Country | United States |