Agilent Technologies Denmark ApS Recall: Brand Name: FLEX Monoclonal Mouse Anti-Human CD20
Agilent Technologies Denmark ApS Recall: Brand Name: FLEX Monoclonal Mouse Anti-Human CD20 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Agilent Technologies Denmark ApS; Product: Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) Model/Catalog
| Recalling firm | Agilent Technologies Denmark ApS |
|---|---|
| Product | Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) Model/Catalog Number: GA60461-2 Software Version: Not applicable Product Description: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) is intended for use in immunohistochemistry together with the Dako Omnis instrument. This antibody labels cells of the B-cell lineage and is a useful aid for the classification of neoplasms of B-cell derivation. Component: Not applicable |
| Reason for recall | Their is the potential for weak staining which may result in false negative CD20 identification. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | US Nationwide distribution including in the states of Alabama, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Nevada, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Penns |
| Recall initiated | 20250811 |
| Recall number | Z-2563-2025 |
| Category | device |
| State | N/A |
| Country | Denmark |