← RecallGuard
HomeRecalls

Alere San Diego, Inc. Recall: The Cholestech LDX¿ System combines enzymatic methodology an

Alere San Diego, Inc. Recall: The Cholestech LDX¿ System combines enzymatic methodology an is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Alere San Diego, Inc.; Product: The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, H

Recalling firmAlere San Diego, Inc.
ProductThe Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987
Reason for recallCholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA
Recall initiated20220624
Recall numberZ-1508-2022
Categorydevice
StateCA
CountryUnited States

🔍 Search all recalls →