Ameditech Inc Recall: DrugSmart Cup 10, Item No. 61093D-3C, 61085D, 61015D
Ameditech Inc Recall: DrugSmart Cup 10, Item No. 61093D-3C, 61085D, 61015D is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Ameditech Inc; Product: DrugSmart Cup 10, Item No. 61093D-3C, 61085D, 61015D; Reason for recall: Ameditechs Drugs of Abuse Tests listed above have shown reduced re
| Recalling firm | Ameditech Inc |
|---|---|
| Product | DrugSmart Cup 10, Item No. 61093D-3C, 61085D, 61015D |
| Reason for recall | Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide. |
| Classification | Class III (least serious — unlikely to cause adverse health consequences) |
| Status | Terminated |
| Distribution | Nationwide Distribution. |
| Recall initiated | 20151116 |
| Recall number | Z-0497-2016 |
| Category | device |
| State | CA |
| Country | United States |