American Contract Systems, Inc. Recall: Pacemaker, COPM11B; Medical convenience kit

American Contract Systems, Inc. Recall: Pacemaker, COPM11B; Medical convenience kit is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: American Contract Systems, Inc.; Product: Pacemaker, COPM11B; Medical convenience kit; Reason for recall: ACS identified that due to an inoperable chart r

Recalling firmAmerican Contract Systems, Inc.
ProductPacemaker, COPM11B; Medical convenience kit
Reason for recallACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS Nationwide distribution in the states of MO, MN, MA, OH, NE.
Recall initiated20240807
Recall numberZ-3244-2024
Categorydevice
StateMO
CountryUnited States