← RecallGuard
HomeRecalls

American Medical Systems, Inc. Recall: AMS 700 Series Inflatable Penile Prosthesis Accesso

American Medical Systems, Inc. Recall: AMS 700 Series Inflatable Penile Prosthesis Accesso is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: American Medical Systems, Inc.; Product: AMS 700 Series Inflatable Penile Prosthesis Accessory Kit (AMS 700 Line Accessory Kit), 72401850. Keith Ne

Recalling firmAmerican Medical Systems, Inc.
ProductAMS 700 Series Inflatable Penile Prosthesis Accessory Kit (AMS 700 Line Accessory Kit), 72401850. Keith Needles are part of the AMS 700 Accessory Kit. In conjunction with the Furlow Insertion Tool (Furlow), the Keith Needles facilitate the passing of the pulling sutures through the glans to place the AM3-700 cylinders in the corpus cavernosa.
Reason for recallAMS' supplier has issued a recall of certain lots of Lightning Bolt Keith needles supplied in AMS 700 Accessory Kits. Based on this, AMS is recalling AMS 700 Accessory Kits which contain these lots of needles. The supplier made the decision to issue this recall after an internal investigation that was requested by AMS. The investigation identified a potential nonconformity that may lead to a bent or broken Keith Needle when used with the Furlow tool.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution US (nationwide) including PR and the states of AL. AZ. AR, CA, CO, CT, IL, IN, KS, KY, MD, MA,MI, MN, MO, NY,NC, OH,OR, TN, TX, VA, and WA, and the countries of ARGENTINA, AUSTRALIA, BRAZIL, CANDA, GERMANY, FRANCE, GREAT BRITAIN, LEBANON, PANAMA, SOLOMON ISLANDS, SWEDEN, and C
Recall initiated20141010
Recall numberZ-0209-2015
Categorydevice
StateMN
CountryUnited States

🔍 Search all recalls →