← RecallGuard
HomeRecalls

American Medical Systems Innovation Center - Silic Recall: Tool, Stripper W/BLD, 18S, Nona

American Medical Systems Innovation Center - Silic Recall: Tool, Stripper W/BLD, 18S, Nona is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: American Medical Systems Innovation Center - Silicon Valley; Product: Tool, Stripper W/BLD, 18S, Nonautoclavable; The Fiber Stripper is a tool prov

Recalling firmAmerican Medical Systems Innovation Center - Silicon Valley
ProductTool, Stripper W/BLD, 18S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.
Reason for recallvalidation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israe
Recall initiated20150528
Recall numberZ-2178-2015
Categorydevice
StateCA
CountryUnited States

🔍 Search all recalls →