← RecallGuard
HomeRecalls

Arkray Factory USA, Inc. Recall: arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, P

Arkray Factory USA, Inc. Recall: arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, P is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Arkray Factory USA, Inc.; Product: arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Bl

Recalling firmArkray Factory USA, Inc.
Productarkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
Reason for recallOne lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusTerminated
DistributionDistributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
Recall initiated20151218
Recall numberZ-0631-2016
Categorydevice
StateMN
CountryUnited States

🔍 Search all recalls →