Atrium Medical Corporation Recall: Express Dry Seal Chest Drains (Express): 4000-100N DRAI
Atrium Medical Corporation Recall: Express Dry Seal Chest Drains (Express): 4000-100N DRAI is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Atrium Medical Corporation; Product: Express Dry Seal Chest Drains (Express): 4000-100N DRAIN, EXPRESS, SINGLE; 4050-100N DRAIN, EXPRESS, BRU; Reas
| Recalling firm | Atrium Medical Corporation |
|---|---|
| Product | Express Dry Seal Chest Drains (Express): 4000-100N DRAIN, EXPRESS, SINGLE; 4050-100N DRAIN, EXPRESS, BRU |
| Reason for recall | Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Brazil, Brunei Darussalam, Canada, Chile, China, Czech Republic Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Korea, Republic of Kuwai |
| Recall initiated | 20230918 |
| Recall number | Z-0318-2024 |
| Category | device |
| State | NH |
| Country | United States |