← RecallGuard
HomeRecalls

Barco N.V. Recall: MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Barco N.V. Recall: MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Barco N.V.; Product: MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display; Reason for recall: We have identified a potential issue wit

Recalling firmBarco N.V.
ProductMNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display
Reason for recallWe have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusCompleted
DistributionUS: FL
Recall initiated20240424
Recall numberZ-2332-2024
Categorydevice
StateN/A
CountryBelgium

🔍 Search all recalls →