Bausch & Lomb Inc Irb Recall: LASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF
Bausch & Lomb Inc Irb Recall: LASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Bausch & Lomb Inc Irb; Product: LASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF/PRODUCT CODE E7549ADB, STERILE, Rx Only Product Usage:
| Recalling firm | Bausch & Lomb Inc Irb |
|---|---|
| Product | LASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF/PRODUCT CODE E7549ADB, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures. |
| Reason for recall | The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | US Nationwide Distribution |
| Recall initiated | 20170303 |
| Recall number | Z-0868-2018 |
| Category | device |
| State | NY |
| Country | United States |
$29/mo
Try RecallGuard →