← RecallGuard
HomeRecalls

Bausch & Lomb Inc Irb Recall: LASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF

Bausch & Lomb Inc Irb Recall: LASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Bausch & Lomb Inc Irb; Product: LASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF/PRODUCT CODE E7549ADB, STERILE, Rx Only Product Usage:

Recalling firmBausch & Lomb Inc Irb
ProductLASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF/PRODUCT CODE E7549ADB, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.
Reason for recallThe complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUS Nationwide Distribution
Recall initiated20170303
Recall numberZ-0868-2018
Categorydevice
StateNY
CountryUnited States

🔍 Search all recalls →