Baxter Healthcare Corporation Recall: L5C4531 X1 Integrated APD Set with Cassette 3-Prong
Baxter Healthcare Corporation Recall: L5C4531 X1 Integrated APD Set with Cassette 3-Prong is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Baxter Healthcare Corporation; Product: L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene
| Recalling firm | Baxter Healthcare Corporation |
|---|---|
| Product | L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide. |
| Reason for recall | Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution. |
| Recall initiated | 20181010 |
| Recall number | Z-0030-2020 |
| Category | device |
| State | IL |
| Country | United States |