Bayer Medical Care, Inc. Recall: Overhead Counterpoise Systems (OCS) OCS CONF an accessory
Bayer Medical Care, Inc. Recall: Overhead Counterpoise Systems (OCS) OCS CONF an accessory is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Bayer Medical Care, Inc.; Product: Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a suppor
| Recalling firm | Bayer Medical Care, Inc. |
|---|---|
| Product | Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288 |
| Reason for recall | The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | The devices were distributed to the following US states: OH, TX, and WV. |
| Recall initiated | 20180628 |
| Recall number | Z-2961-2018 |
| Category | device |
| State | PA |
| Country | United States |