BD SWITZERLAND SARL Recall: BD Alaris PCEA Administration Set, Microbore Tubing with Yello

BD SWITZERLAND SARL Recall: BD Alaris PCEA Administration Set, Microbore Tubing with Yello is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: BD SWITZERLAND SARL; Product: BD Alaris PCEA Administration Set, Microbore Tubing with Yellow Identification Stripe ASV, Catalog Number 30893-04 (O

Recalling firmBD SWITZERLAND SARL
ProductBD Alaris PCEA Administration Set, Microbore Tubing with Yellow Identification Stripe ASV, Catalog Number 30893-04 (OUS Only) - Product Usage: Microbore Tubing with Yellow Identification Stripe Anti-siphon Valve Use Aseptic Technique Set cannot be used for gravity infusion. 1. Connect PCA syringe to female luer on administration set. 2. Push on syringe plunger to prime set. 3. Refer to PCA module Directions for Use for complete loading and operating instructions. Notes: The set should be changed according to facility protocol or in accordance with currently recognized guidelines for IV therapy
Reason for recallThe products have the potential to leak between the connection of the male luer and the yellow striped tubing. This leak may be observed while priming or during use. The most likely health consequence associated with a leak in an administration set would be inadequate pain control for the patient due to an under-infusion of medication.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide distribution - US Nationwide Distribution and the countries of Canada, Belgium.
Recall initiated20200409
Recall numberZ-1969-2020
Categorydevice
StateN/A
CountrySwitzerland